Series 34 Sanitary Dio-Seal Isolator Installation
|
The sanitary isolator is designed to mate with the
Tri-Clamp pipe sections. Flange gaskets are not used
since the Sensor Ring contains integrally molded Tri-Clamp
gaskets. Use appropriate sanitary clamps to secure isolator
in pipeline, applying normal tightening torque to assure a
leak-free installation. This integral gasket design facilitates
factory calibration and ensures accuracy in field installation.
The annular volume between isolator body and Sensor Ring and
the instrument internal volume are filled with a food grade
Sensor Fluid, providing the hydraulic link required to transfer
the process pressure signal to the instrument. The Sensor Fluid
must be compatible with the Sensor Ring material and the process
fluid.
Working pressure is nominally the Tri-Clamp pressure rating at the
process operating temperature. The isolator can be operated up to the
maximum allowable temperature rating of the Sensor Ring material.
For verification of allowable temperature ratings, consult Factory.
|
To print out this and other important information, click |
|